Aluro Aerosol Max Manuale di istruzioni - Pagina 7

Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature mediche Aluro Aerosol Max. Aluro Aerosol Max 8. Extra oral suction unit

CLEANING & DISINFECTING
Hood housing piece
Hood cover
Machine body
CLEAN DOWN THE AEROSOL MAX ON A PER PATIENT BASIS;
Step 1 - Spray the hood cover with hydrogen peroxide 3% inside and outside and leave for 15
minutes before wiping down. After 15 minutes, wipe with a cloth.
Step 2 - Wipe the machine body and handles using a cleaning and disinfection wipe to clean
and disinfect the product surface and handles, paying particular attention to the control
panel.
Step 3 – Wipe along the outside of the flexible suction tubing with a cleaning and
disinfection wipe to clean and disinfect the outer of the flexible tubing.
Step 4 – Turn the machine onto speed 2-3 and spray 3% hydrogen peroxide into the hood
housing piece. Use 7-10 sprays, sprayed into the hood housing piece of the machine. Let the
unit run for 3 minutes, then turn off.
Step 5 – Wipe the hood housing piece with a cleaning and disinfection cloth after each
patient.
Aluro recommends to use non-alcohol cleaning & disinfection wipes like ECOCLOTH,
or another neutral detergent and non-alcohol disinfection wipes. You can purchase 3%
Hydrogen Peroxide Spray from Aluro HealthcareNZ.
Flexible suction
tubing
Lifting handles
Control panel
www.aluroequipment.co.nz //
7
PH 0800 800 805