Troubleshooting

Faults on lifters that do not function as they should
after assembly (page 5) or after the pre-use check
has been completed (page 7), can be localized us-
ing the troubleshooting table below.
Symptom
Possible cause/action
The lifting column
The lifting column is not correctly
is wobbly
positioned in the chassis or the
locking handle is not fully locked
/ verify position of lifting col-
umn and locking handle acc. to
mounting instructions on page 5.
The lifting column can not be
correctly fitted / faulty safety
mechanism in chassis, contact
local representative for service
The lifter moves
The castors move unevenly be-
unevenly on a
cause of pile and dust / clean
plane surface
castors as described on page
16.
Fault in the leg spreading mech-
anism / contact local representa-
tive for service.
The lifter only
Hand control has failed / test
moves in one di-
with another hand control and
rection - either up
change if defective.
or down
The control electronics has failed
/ contact local representative for
service and use manually emer-
gency lowering, where neces-
sary
The lifter runs out
Short circuit between hand con-
of control
trol and lifter / clean hand control
plug surfaced with alcohol to re-
move grease
Defective hand control / replace
hand control / contact local rep-
resentative for service
The control electronics has failed
/ contact local representative for
service
BM09201 Molift SMART 150 English - Rev E 10/2012
Symptom
Possible cause/action
The lifting arm (lift-
Emergency stop button is acti-
ing motor) does
vated / turn button clocwise to
not move
reset emergency stop, see p. 9
Battery is empty / change to a
new battery or recharge battery,
see page 10
The control electronics of the lift-
er is overheated / Leave lifter to
cool down, see page 10
The lifter is overloaded / make
sure the load does not exceed
150 kg, see page 9
The hand control cable is un-
plugged / reconnect plug
Hand control has failed / replace
hand control / contact local rep-
resentative for service
Defective control electronics /
contact local representative for
service
Battery does not
Faulty battery / try another bat-
recharge
tery / replace battery
Faulty battery charger / try an-
other charger / contact local rep-
resentative for service
If you have access to more than one lifter, it may
be helpful to change parts across lifters to help
identify any faults. For instance, if you suspect the
fault to be inside the hand control, exchange the
hand control with at functional one from another
similar lifter to see if the problem is fixed.
If you are not able to solve the fault or problems
by yourself using this manual, please contact lo-
cal authorised service representative. If you do
not know whom to contact locally, please contact
your dealer or Etac AS to get help.
Page 15 of 20