Your new Aseptico Express Air Portable Dental System is the
finest portable electric dental system available. The ADU-17A
features a handpiece control, 3-way air/water syringe, HVE and
saliva ejector vacuum systems, a self contained water system,
oilless air compressor, and vacuum pump. The dual voltage
system comes in a case with wheels and a handle for maximum
portability.
Congratulations!
This system is engineered to provide many years of reliable
service. Please read the instructions provided in this manual
to receive the best and longest service from your Aseptico
equipment.
Separate manuals may be provided to cover the operation and
maintenance of other accessories for your unit.

PACKAGE CONTENTS:

• Portable Case with Compressor and Vacuum Pump
• TA-90D 3-Way Air/Water Syringe
• HVE and Saliva Ejector Vacuum Hoses with Valves
• Water Supply Bottle
• AA-42 Air Foot Control (Complete Assembly with Fittings: PN
330699)
• Power Cord
• Operation & Maintenance Manual
• Highspeed or Lowspeed Pneumatic Handpiece (Purchased
separately)
• AWS-1C Cavitron Connector
• Two Spare 6A Fuses for 120V Operation

SAFETY PRECAUTIONS:

Aseptico accepts no liability for direct or consequential injury or
damage resulting from improper use, arising in particular through
the non-observance of the operating instructions, or improper
preparation and maintenance.
WARNING: Sterilize before first and all uses. Clean, disinfect,
and sterilize new or repaired instruments before first
use. Only use sterilized handpieces and instruments
during treatment. Non-sterile handpieces and instruments
may cause bacterial or viral infections. Always sterilize
instruments after operation.
CAUTION: Always examine unit components for damage before
commencing treatment. Damaged components must not be
used and must be replaced.
CAUTION: For device transport, always keeps the pull handle
in hand during transport and ensure stable platform before
releasing.
WARNING: Use for intended purposes only. Failure to observe
the operating instructions may result in the patient or user
suffering serious injury or the the product being damaged,
possibly beyond repair. Before using this product, make
sure that you have studied and understood the operating
instructions.
WARNING: For use by qualified and trained personnel only.
WARNING: Do not install where there is a risk of an explosion.
The Express Air is not intended for operation in the presence
of flammable anesthetics or gases.
WARNING: Always operate a high-speed handpiece with water
coolant. Operating a high-speed handpiece without water
coolant can cause thermal injury to the patient and damage
the handpiece.
WARNING: Only connect device to AC outlet with Protective
Earth (Safety Ground) connection.
WARNING: The power supply cord is the mains appliance
disconnect means. Before use, position the equipment so it
is readily accessible.
Expected Service Life:
The expected service life is seven years from the date of first use.
3.