The dispenser will not
activate despite working
batteries and a Clean
Cassette is installed with
disk covers present, no light
indicators are present
Alternating yellow, green,
and/or red cassette and
battery indicator lights
Solid red light ring
Solid Red Clean Cassette
indicator light (briefly
illuminated)
12
Perform system reboot:
1. Remove Clean Cassette and batteries.
2. Wait 5 seconds.
3. Reinstall batteries.
If green battery indicator light appears, ensure
shipping insert is removed and reinstall Clean
Cassette.
If no green Clean Cassette indicator light
appears, contact Customer Service.
Perform system reboot:
1. Remove Clean Cassette and batteries.
2. Wait 5 seconds.
3. Reinstall batteries.
If green battery indicator light appears, reinstall
Clean Cassette.
Disk cover may be jammed in Clean Cassette.
1. Remove Clean Cassette.
2. Remove any visibly jammed disk covers
from the backing tape. If no jammed disk
covers
are visible, perform system reboot
(described above).
3. Reinstall Clean Cassette
If light ring again turns red, replace with new
Clean Cassette.
Clean Cassette failed authentication.
Check to ensure the shipping insert has
been removed.
Perform system reboot:
1. Remove Clean Cassette and batteries.
2. Wait 5 seconds.
3. Reinstall batteries.
If green battery indicator light appears, reinstall
Clean Cassette.
If no green Clean Cassette indicator light
appears, contact Customer Service.