Sfoglia online o scarica il pdf Manuale di istruzioni per l'installazione e la manutenzione per Strumenti di misura Ashcroft AM. Ashcroft AM 19. Bimetal thermometer
2.5 Signs/safety marking
The device is provided with a label. The label shows the type designation, measuring range, serial number, year of
manufacture, certificate of approval number, filling medium, Ex marking (including X for special conditions of use) and
manufacturer.
On the dial the manufacturer, measuring unit, accuracy class, note Silicone-free and the applied standard are visible.
The device may have other labels and safety marks indicating special conditions of use:
Hot surface
Note on calibration
The operator must check the label, which is important for the use in potentially explosive atmospheres, at regular
intervals to ensure that it remains legible.
The outer packaging is labelled with the type designation, order number, item number, measuring range and
manufacturer's data.
2.6 Safety devices
The window is made of laminated safety glass, no other material options are permitted here.
2.7 Environmental protection
This device may optionally contain filling liquid (silicone oil). The provisions of the Regulation on Registration, Evaluation,
Authorization and Restriction of Chemicals (REACH) 2006/1907/EC must be observed, the corresponding safety data
sheets of the manufacturers of the chemicals, are available for download on our website. At the end of the product life
cycle we recommend to recycle the devices, as they are mostly made of stainless steel. Instructions for disassembly,
material separation and disposal can be found in chapter 11
3 Use in potentially explosive atmospheres according to Directive
2014/34/EU ATEX & Regulation SI 2016, No. 1107
3.1 Range of use:
Potentially explosive atmospheres Zone 1 and 2, as well as 21 and 22, hazard due to gases and dry dusts.
The permissible environment has the usual oxygen content (21%), ambient pressure 80 kPa (0.8 bar) to 110 kPa (1.1
bar).
The requirements of the applicable standards EN ISO 80079-36, EN ISO 80079-37, EN 60079-0 were considered by
an ignition hazard assessment. The applicable requirements of these standards have been met.
For the European Union the documentation has been filed with TÜV-Nord-Cert NB 0044 (see Declaration of
Conformity). For the United Kingdom, the documentation has been filed with Element Materials Technology AB 0891
(see Declaration of Conformity).
Ambient temperatures:
Execution
Dry finish
Silicone filled devices
Min. / Max. permissible ambient temperature
-20°C to +60°C