Installation and Maintenance Instruction Manual for All
Stainless Steel Gas Actuated Thermometer Model S5500
(non-electrical device) in a configuration for explosion risk areas pursuant to
Directive 2014/34/EU (ATEX)Zone I and II, Category 2, Group II for gas and dust
6.3.2 Capillary line
The capillary tube is located in a protected position within
the stem, or is executed as a flexible part with or without
additional flexible armor or a plastic sheathing.
6.3.3 Case
The window uses multi-layer safety glass. This device is
fitted with a rear plug capable of being blown out.
6.3.4 Dial with pointer
The gauge is equipped with a dial and pointer pursuant to
EN 13190.
6.3.5 Instrument connection
The instrument connection is located on the underside of
the gauge and can be executed as a threaded or flanged
connector. Union instrument connections permit the case
to be oriented according to the point of installation.
6.4 Accessories
Please contact the manufacturer regarding special tools
and accessories
7 TRANSPORT
7.1 Safety
The thermometer should be protected against the effects
of knocks and impacts. The device should only be
transported in the packaging provided, to protect against
glass breakage. The device should only be transported in
a clean condition (free from residues of measuring media)
7.2 Transport inspection
The delivery is to be checked for completeness and
damage during transport. In the event of damage during
transport, the delivery is not to be accepted, or only
accepted subject to reservation of the scope of the
damage being recorded and, if necessary, a complaint
initiated.
7.3 Storage
The gauge is to be stored in dry, clean conditions, within
a temperature range of 20 to +60 °C, protected against
direct exposure to sunlight and protected against impact
damage.
© 2021 Ashcroft Inc., 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0499, www.ashcroft.com
All sales subject to standard terms and conditions of sale. S5500_Gas_Act_Therm_RevA_(008-10293)_05-26-21
8 MOUNTING/INSTALLATION
8.1 Safety
Thermowells must be used
for all processes and flowing
media subjected to pressure.
They protect the thermometer
stem against corrosion and
mechanical damage, and
permit the thermometer to be
removed without leaks.
8.2 Preparations
(requirements for the
installation location)
• A check on suitability of the
device and of the thermowell
that may be required for the
medium to be measured, the
arrangements in the scope
of measurement range and the extent of the protection
against special conditions such as vibrations.
• A bracket must be installed to support the gauge if the
metering pipe is not able to provide adequate support
• Devices with a blow-out require a minimum spacing to
the rear (20 mm).
• The operating temperature range is -30°C to + 80 °C.
8.3 Moutning/Intallation
• Using appropriate accessories, the device can be
installed on flat walls, mounting plates, on pipes or in
panels or boards.
• Ex works, the device is supplied and calibrated for
vertical installation.
• If the installation location deviates from the vertical (max.
± 5°), the zero setting on the indicator must be corrected
(see 7.4.1 Zero point adjustment).
8.3.1 Every angle connection (optional)
Only adjust the every angle connection if this is necessary
during fitting or removal.
• Use the every angle connection to place the sensor in the
correct position prior to installation, proceeding as follows:
• Position the thermometer housing in a straight alignment
(Position "C"
• Using the screws marked "A", loosen until the joint can
be turned freely through 180° on the lower part of the
housing and the sensor.
4