INSTALLATION/USE

1.
Unroll the mattress.
2.
Position it on the base of the bed, ensuring the cushion "head" and "foot" symbols are placed correctly.
3.
Hook the compressor to the bed frame (the foot).
4.
Guide the power lead for the compressor along the bed to the nearest power point and connect it.
N.B.: If you have a high-speed inflation pump, after stage 4,
-
Connect the power lead for the inflation pump to the mains,
-
Blank off the connection to the mattress feed pipe with the attached plug.
-
Set the CPR valve to "OPEN",
-
Connect the high speed inflation pump as shown in the diagram below,
(Comment: the pump inflates in one direction and deflates in the other),
-
Press the start button until the mattress appears to be fully inflated (approximately 2 minutes)
5.
Close the mattress CPR valve (set the dial to "CLOSE")
6.
Connect the mattress feed pipes to the compressor connector
7.
Press the Start/Stop button. The lights come on for 1 second and a BEEP sounds.
8.
Wait until the "Wait" indicator stops flashing, indicating the end of the auto-inflation phase. (Without high speed inflation, this
step lasts about 20 minutes for a mattress 87 cm wide and 30 minutes for a mattress 117 cm wide)
9.
Lay the patient out flat on the mattress.
10.
Start a pressure calculation by pressing the
automatically after a 5 minute pause.
During the calculation, the pressure button indicator stays on and the pressure indicators come on alternately. (For a precise
calculation, the patient should not move excessively). The calculation lasts a maximum of 6 minutes.
When the calculation is finished, the pressure button indicator goes out and the pressure indicators are lit continuously to
indicate that the pressure is now matched to the patient's morphology.
The compressor then starts DYNAMIC mode automatically.
11.
Selecting the position corresponding to that of the patient. (see § CONTROL PANEL)
COMMENTS:
Unlock the keypad before making selections by holding the button down for a little while (Example: to change mode)
To recalculate the inflation pressure, press again on the pressure button and the pressure indicators will come on alternately.
If installation was completed without the patient present, you must rerun the calculation once the patient is lying
on the mattress.
It is possible, exceptionally, to modify the setting obtained by automatic calculation by holding the "pressure" button down.
The principal information for using the compressor is printed on its side.
After 30 minutes has passed without selection, the intensity of the indicators reduces in order not to inconvenience the patient
during the night. Pressing a key resets maximum intensity for 30 minutes.
WARNING: for a precise pressure calculation, the patient should be lying flat on the bed and should not make any sudden
movements.
WARNING:
the power lead should be should be installed so as to avoid any interference with the articulated parts or the wheels of the bed and
employees cannot trip over it. There is a risk of physical injury and damage to the equipment. Use the removable hooking systems
supplied.
button while the LED is flashing. If you do nothing, the calculation will start
- 8 -