- ページ 14

計測機器 Flomotion Systems LT260のPDF 操作およびメンテナンスマニュアルをオンラインで閲覧またはダウンロードできます。Flomotion Systems LT260 18 ページ。 Ultrasonic level transmitter

[P10] Dampening

[Default: 001.00]
Sets the display value response time.
Option
000.01
Slowest
100.00
Fastest

[P11] Detection Threshold Voltage

[Default: 03 and is acceptable for most applications]
Sets the detection of the returned echo. For noisy applications increase the detect threshold. For
detection of weak signals, decrease the detect threshold value.
Range: 03 (Low) - 15 (High)

[P12] Output Power

[Default: 2]
Option
1
Low Power
2
Normal Power
3
High Power
4
Maximum Power
Description
Slowest response to level change.
Minimum: 000.01 - Maximum: 100.00
Fastest response to level change.
Level
Description
Use for short range applications
Use in normal conditions
Use in long range measurement
Use in dusty, steamy or turbulent applications.
14