- ページ 10

医療機器 AMD RCS-100のPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。AMD RCS-100 12 ページ。 Multipurpose camera & scope system

AMD RCS-100 ユーザーマニュアル
EMC (electromagnetic compatibility)
• During installation and operation of the device, observe the
following instructions:
• Do not use the device simultaneously with other electronic
equipment to avoid electromagnetic interference with the
operation of the device.
• Do not use or stack the device near, on, or under other
electronic equipment to avoid electromagnetic interference
with the operation of the device.
• Do not use the device in the same room as other electronic
equipment, such as life-support equipment that has major
effects on the life of the patient and results of treatment,
or any other measurement or treatment equipment that
involves small electric current.
• Do not use cables or accessories that are not specified
for the device because that may increase the emission of
electromagnetic waves from the device and decrease the
immunity of the device to electromagnetic disturbance.
• Do not touch the pins connecting the control unit to the
lenses or the signal pad on the lenses without special pre-
cautions.
Attention:
Medical electrical equipment (ME) is subject to special pre-
cautions regarding electromagnetic compatibility (EMC).
Portable and mobile radio frequency communication devices
can affect medical electrical equipment. The ME device is for
operation in an electromagnetic environment of home health
care and intended for professional facilities such as industri-
al areas and hospitals.
The user of the device should ensure that it is operated within
such an environment.
Warning:
The ME device may not be stacked, arranged or used directly
next to or with other devices. When operation is required to
be close to or stacked with other devices, the ME device and
the other ME devices must be observed in order to verify pro-
per operation within this arrangement. This ME device is only
intended for use by medical professionals. This device may
cause harmful interference or interfere with the operation of
nearby devices. It may become necessary to take appropriate
measures, such as redirecting or re arranging the ME device
or shield.
The ME device assessed does not exhibit any essential per-
formance features in the sense of EN60601-1, which would
present an unacceptable risk to patients, operators or third
parties in the event that it should present an outage or the
power supply should disconnect.
Warning:
Portable RF communications equipment (radios) including
accessories, such as antenna cables and external antennas,
should not be used in closer proximity to modules than spe-
cified by the RCS-100 parts and lines manufacturer 30 cm
(12 inch). Failure to comply may result in a reduction in the
device's performance features.
22