13. Specifications
■ CO2 & Temperature specification:
CO2 Specification
Measurement Range
Display Resolution
Accuracy
Pressure Dependence
Response Time
CO2 AL1 (TWA )
CO2 AL2
CO2 AL3
Sound Alarm
Warm-Up Time
Temperature Specification:
Temperature Range
Display Resolution
Display Options
Accuracy
Response Time
Operating Conditions:
Temperature
Humidity Range
Storage Conditions:
Storage Temperature
■ Power Supply
DC
Power
Supply
AC adapter
Battery
Voltage
■ Relay output
Relay 1
Relay 2
Relay 3
The Peak current for all relays is less than 2A@30 VDC or 250 VAC, SPDT.
■ Weight and Dimension
Weight
SEU
Dimensions (LxWxD)
Weight
RDU
Dimensions (LxWxD)
0 - 50,000ppm (5%) display
10ppm at 0~10,000ppm;
100ppm at 10,001~50,000ppM
+/-200ppm or +/-10% reading
Auto pressure compensation, built in barometer. (50 to 110 kPa)
CO2: <2min by 90%
5000ppm fixed
5000, 1.0%, 1.5%, 2.0%, 2.5%, 3.0%. Default AL2= 1.5%.
2.0%,2.5%,3.0%,3.5%,4.0%. Default AL3= 3.0%
80db@10cm
<60 seconds at 22°C
0°C to 50°C (32°F to 122°F)
0.1°C (0.1°F)
°C/°F
±1.5°C(±2.7°F) when CO2 level is below first alarm level
20-30 minutes (case must equalize with environment)
0°C to 50°C ( 32°F to 122°F)
0 ~ 95% RH non-condensing
-20°C to 60 °C (-4°F to 140°F)
9~32VDC (12~32VDC recommended), 2A.
Input: 100~240 VAC,50/60Hz, 0.6A
Output: 12VDC, 2000mA.
6VDC (5.4V~7.0V),
recommended capacity is 12AH
Operates at CO2 TWA
Operates at Alarm 2 level for CO2
Operates at Alarm 3 level for CO2
1 lb
6.69 x 4.96 x 2.73 inches
0.4 lb
5.51 x 3.54 x 1.89 inches