Connecting the electronic measuring devices
CAUTION
The permissible overpressure must not be exceeded, not even in case of pressure peaks!
Install protection elements (e. g. damping screw)
The electronic measuring devices must be mounted at the bottom fitting with a size 19 wrench
(G ¼) or size 32 (G ½) and tightened to a torque of 45 Nm. The installation position is arbitrary.
CAUTION
Avoid electrostatic discharges!
Ground the housing
Electrical connection of the individual models (see type label) must be carried out according to the
chart below. The maximum power supply values are stated on the type label.
Current output
UPA 2
4... 20 mA
UTA 2
(two-wire system)
Connection
Plug
Cable
Supply (+)
1
brown
Supply (-)
2
white
Signal (+)
---
Signal (-)
---
Connection chart for temperature sensor UTF 2
PT100
4
Voltage output
0... 10 V DC
(three-wire system)
GL cable
M12
Plug
Cable
GL cable
white 1
1
1
brown
blue 1
2
2
blue
3
white
---
Plug M12 x 1, 4-pin
6
M aintenance/Cleaning
5 B 5 B 5 B 5 B
The electronic measuring devices are calibrated in the factory and require not maintenance.
If faults occur which cannot be remedied please contact our local representative closest to you.
7
T echnical Data
6 B 6 B 6 B 6 B
see type label
Dimensions
M12
blue 1
1
blue 2
3
white 1
2
IMPORTANT
Electronic pressure transducer UPA 2 (G1/4")
in mm (inch) –
Process connection G1/4 acc. to DIN ISO 228-1
Profile seal acc. to DIN 3869
Cable gland Pg 11, clamping range Ø 6...8 mm
5