- ページ 14
医療機器 Daavlin DT Home-ControlのPDF 操作マニュアルをオンラインで閲覧またはダウンロードできます。Daavlin DT Home-Control 19 ページ。 Modes of operation - timed mode
DO NOT REPLACE LAMPS INDIVIDUALLY. ALL LAMPS SHOULD BE REPLACED AT THE
SAME TIME TO AVOID UNEVEN TREATMENTS OR ERYTHEMA. PLEASE CONTACT THE
DAAVLIN SERVICE DEPARTMENT PRIOR TO REPLACING THE LAMPS TO OBTAIN
IMPORTANT INFORMATION ON HOW THE OUTPUT OF THE UNIT WILL BE AFFECTED BY
THE CHANGE.
Because of differences between brands, always replace lamps with the same brand as originally
installed. Consult the 6.1
and contact the Daavlin service department for a lamp replacement quote. They can be reached
by dialing 1-800-322-8546 or via email at [email protected].
15.3.1 Lamp Replacement
1. First, unplug the device, then, using a Phillips head screwdriver, remove the two screws
located on each side of the device. Then,
2. Lift off the top of the device to expose the lamps.
3. On each lamp there are two clear plastic clasps that hold the lamps in position. To remove
them position both thumbs on the top of the clasp and use your index fingers to pull the
sides of the clasp outward until it releases, and the clasp can be removed.
4. On top of each lamp socket there will be a red button that, when pushed, will allow you to
remove the base of the lamp from its corresponding lamp socket. With one hand firmly
press and hold the red button down. While holding the red button down use your other
hand to carefully lift the lamp up and out of the socket.
5.
With one hand near the base of the lamp and the other at the opposite end, grasp the
lamp and pull both ends upward at the same time until it clears both clasps and the lamp
holder/socket.
6. To replace the lamps in their lamp-holders, press the base of the lamp straight down. A
distinctive snapping sound will be heard which indicates that the lamp is properly seated.
7. Once reassembled, reset the lamp hours to zero.
15.3.2 Resetting Lamp Hours
When changing lamps, it is important to reset the lamp age to zero so the new lamp's
operational hours can be tracked. Please contact the Daavlin service department for
instructions on resetting the device's lamp hours.
16.0 Environmental Specifications
The 1 Series should be used in an electromagnetic environment as described below.
Emissions Test
RF Emission Following
CISPR 11 (EN 55011)
MNL-00049 [1]
Lamp Specification Guide
Table 3 Electromagnetic Emissions
Conformity
Group 1
Proprietary and Confidential. Printed copies will not be updated.
on page 4 for replacement lamp guidance
EMC Environment Guide
Test unit only radiates RF energy for
internal use in powering lamps, and it
seems unlikely that nearby medical
devices would be affected.
Page 14 of 19