アクセサリー EKO HF-10SのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。EKO HF-10S 20 ページ。 Heat flow sensor

2. Traceability
Traceability of EKO Heat Flow Sensor is as described below:
Guarded Hot Plate Method
(GHP Method)*
EKO Reference Plate
(Silicon Rubber)
EKO Reference
Heat Flow Sensor
EKO Heat Flow Sensor
MF-180, MF-180M
HF-30S, HF-10S
Figure 6-1. Traceability
*The Guarded Hot Plate Method (GHP Method) is a thermal conductivity measurement method of insulation
material by the absolute method based on JIS A 1412. By measuring the power applied to the main heat plate,
calculate the heat flow which flows through the sample piece and determine the heat flow rate from the
temperature difference and thickness of the sample.
By keeping the protecting plate around the main heat plate at same temperature as the main heat plate, the
heat flow runs through the sample achieves the primary flow (the heat flows in perpendicular to the sample).

6-3. Troubleshooting

Check the following items in case of trouble with the instrument. If any questions should remain, contact EKO
for further technical support.
Table 6-1. Troubleshooting
Failure
There is no output.
Output is very low.
EKO INSTRUEMTNS CO., LTD. Heat Flow Sensors Instruction Manual Ver.7
JIS A 1412
Check the resistance.
Check the connection of the lead wires.
Check the sensor contacting and adhesion conditions.
Calibrate EKO Reference Plate with the device
which complies with GHP method.
Using calibrated EKO Reference Plate, calibrate the
reference heat flow sensor periodically.
Calibrate the production heat flow sensor according to
the HFM method complying with JIS A 1412.
Action
Pg. 13