C800 Analyser Operating Manual
RANGE – Select the range appropriate for the
process. Lower ranges give greater accuracy
and higher ranges permit higher readings.
DSL-DSH – This allows the user to set the low
and high process values that correspond to 4
and 20mA respectively. Initially the arrows will
adjust DSL; to adjust DSH press
.
ATC – This allows the user to enable or disable
the automatic temperature compensator and
calibrate the temperature probe. See
section
8.5
for further information. By default, ATC is
enabled.
TEMP COMPENSATION – Conductivity values
vary with changes in temperature. This value
will compensate for the changes of
temperature.
DAMPING – This "dampens" the primary value
by allowing only half the detected
change per time interval (in seconds). Damping
is adjustable between 0 and 10 seconds in
intervals of 0.1. By default, this is set to 0.2
seconds.
RESET RUNTIME – This will display the current
runtime and allows the user to reset the
runtime of the analyser.
902857-A - C800 Analyser Operating Manual.docx
15