バスタブ Astor-Bannerman KivaのPDF 取扱説明書と取付説明書をオンラインで閲覧またはダウンロードできます。Astor-Bannerman Kiva 17 ページ。 Height adjustable bath

Operating the Spa system (optional)
Fill the tub with water. The spa unit is controlled using the round button attached to the bath tub.
To turn the air pump on, pressing the button once will operate the spa at its lowest setting.
Pressing the button for a second time will operate the spa at its medium setting. Pressing the
button for a third time will operate the spa at its highest setting. Once the unit has been turned on,
it will remain active for 20 minutes before automatically turning off. Pressing the button again will
operate the spa system for a further 20 minutes. Use of the spa bath system for longer than 20
minutes is not recommended.
Hygiene management:
The spa system is designed to be self-maintaining with regards to hygiene. Approximately 20
minutes after use, the spa system will automatically turn itself on and operate for approximately 20
seconds. This is called the "Auto-Purge" cycle. Any water remaining within the jet system will be
ejected, keeping the jets clean. When cleaning the bath, using the bath for showering only, or
bathing without the spa system, it is recommended that a manual purge is performed. The Auto-
Purge will then operate 20 minutes later.
Cleaning the bath:
The bath may be cleaned using most mild cleaning agents. Chemical agents and solvents must not
be used, i.e. drain cleaners, acetone, oven cleaners or acids. The taps, side & end panels and
handset should be cleaned using a warm soapy solution and wipe down with a damp cloth.
Attention should always be paid to instruction labels for the cleaning agent.
Troubleshooting:
In the unlikely event that problems occur with theunit, carry out the following easy checks to
determine where or not an engineer is required.
Fault: The Rio or Kiva bath doesn't travel up or down or moves downwards very slowly whilst
beeping.
Cause: The mains power supply has been disconnected for some time and the unit's batteries
are now flat.
Remedy: Examine the mains fuses/circuit breakers and reset as necessary and ensure that the
mains voltage is fully reconnected. Allow the battery back-up system to re-charge for
approximately 2 hours.
Fault: The bath doesnt move upward or downards at all.
Cause: The handset has been pulled from the socket fitted to the underside of the unit.
Remedy: Push the plug fitted to the handset more firmly into the socket.
Fault: Only one end of the bath moves up and down.
Remedy: You need to contact Astor Bannerman as there is a malfunction with the equipment.
Fault: Water temperature is incorrect from thermostatic mixer tap.
Remedy: The thermostatic mixer taps will need recalibrating.
Fault: The water has stopped running.
Remedy: Check that you have both a hot and cold water supply.
Page 9
Version 8 12/2018