水ポンプ AmeriWater SCALA2のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。AmeriWater SCALA2 14 ページ。 Booster pump
COMPONENT IDENTIFICATION
2
COMPONENT IDENTIFICATION
The pressure regulator is designed to reduce incoming water pressure. This regulator has a built-in
thermal expansion bypass feature. This feature prevents downstream pressure from rising to more than
10 psi above the supply pressure.
The pressure regulator has been factory preset to 40 psi.
"DO NOT ADJUST PRESSURE". Failure to comply using a higher pressure will result in damage to the
pump.
2.1 Connections
Models
0080701
0080702
2.2 REPLACEMENT PARTS
3
Inlet Connection
1 ¼" Hose Barb
1 ¼" Hose Barb
SCALA2 BOOSTER PUMP
Outlet Connection
1" Hose Barb
1" Hose Barb
98-2039 Rev A