4. Operating flow chart
Install and connection
-- Press
Unlock mode
-- Set data(high, low, F.S, DP, CAL),refer to 6
DATA Input
DIP S/W setting
--sensor output data ---DIP S/W 1, 2, 3, 4
filter frequency ---- DIP S/W 6, 7, 8
-- Sensor : no load
Auto zero
Press
---Execute auto calibration.
AUTO CAL
Mode Check
---Check function(Relay output,analog output, reset, hold, peak).
Lock mode setting
5. HARDWARE
1) Selection of AC input power
(1) Separate P.C.B from case.
(2) The following label is existed under power transformer
① ② ③ ④
110V : connect 1 and 2
110 220 110
220V : connect 2 and 3
2) Transformation of SAMPLING TIME
15 times/sec (standard).
(1) The following label is existed on the P.C.B
15 times/sec : connect 1 and C
○ ○ ○
15 C 30 30 times/sec : connect 2 and C
--Mount indicator to panel and make connection. (Check AC
POWER 110/220V)
+ key(data protection unlock)
analog output ---- DIP S/W 5
+ key (auto zero)
--Press
+ key(data protection mode)
connect 3 and 4