Gauge Setup Map
Backlight Timer
To Clear Peaks or Line Pressure
Troubleshooting
Err 1
The M1 checks the internal calibration every time it's turned on.
Contact factory for further instructions.
Err 2
The M1 has tried to display a number too large
for the display due to electrical malfunction or numerical error.
Contact factory for further instructions.
Err 5
The M1 pressure sensor is exhibiting behavior out of normal
operating condition. Contact factory for sensor replacement.
1.800.561.8187
continued
( )
next
3x
Lt: 15
(light stays on for 15 seconds)
Lt: 60
(light stays on for 60 seconds)
Lt: On
(light does not turn o automatically)
displayed:
displayed:
or
Err 6
displayed:
Note:
If there is no gauge activity for 10 seconds, the gauge
will return to operating mode. Any Setup options that you have
selected will be enabled.
c r y s t a l e n g i n e e r i n g . n e t
www.
Power Management
*
This menu can be accessed only when the gauge is o .
1
sec
Ultra Low Power Mode
rd
0.1. . . . . . . (1/10 minutes . . . 30,000 hours)
rd
0.2. . . . . . . (1/ 5 minutes . . . . 25,000 hours)
setup
rd
1.0. . . . . . . (1/1 minute . . . . . 10,000 hours)
rd
2.0. . . . . . . (1/ 30 seconds. . . 6,000 hours)
rd
4.0. . . . . . . (1/15 seconds . . . 3,000 hours)
*
O - -
**
O 1H
*
At "Power On, " this setting is displayed as
*
*
At "Power On, " this setting is displayed as
.com
m1
Quick Reference
*
+
( )
next
. . . . . (4/ 1 second . . . . . 600 hours)
. . . (4/ 1 second . . . . . 600 hours)
No Auto o
Auto o 1H
Note:
Whenever one of the rd settings is selected,
the battery icon will change from
to
indicating that the M1 is in Ultra Low Power Mode.
1
=
+
=