- ページ 11

体重計 Doran DS4500のPDF テクニカル・マニュアルをオンラインで閲覧またはダウンロードできます。Doran DS4500 16 ページ。 Infant lactation scale

Doran DS4500 テクニカル・マニュアル

Section 4. Parameter Setup

The DS4500 offers the ability to change the shutoff timer settings. Parameter settings
shown in bold are the default settings.

Entering and Exiting Setup Mode

To enter setup mode from the front panel, press and hold REWEIGH and then
press and hold ZERO. Release both buttons when the display reads CAl 25 if
the scale is in lb units and display CAL 10 if the scale is in kg units.
To exit the setup mode, press UNIT until the display returns to the weigh mode.

Setup Mode Navigation

1. Press UNIT to change the current parameter.
2. Press FEED to change the parameter setting.

Calibration Weight Selection

Cal
25 (10 kg)
45 (20 kg)

Calibration Mode Selections

CAL
0
CAL
25
45
9
25 LB (10 kg)
45 LB (20 kg)
Zero Calibration Point
Press ZERO to calibrate zero point
NOTE: This parameter is only visible when
in the calibration mode.
Span Calibration Point
25 LB
45 LB
NOTE: This parameter is only visible after a
zero calibration is performed.