体重計 Doran DS8030のPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Doran DS8030 19 ページ。 Wheelchair scale
Command (RXD)
Computer to Scale
Z or z
<cr> or W
U
H
R or r
Command (RXD)
Computer to Scale
DR<cr><lf>
Command (RXD)
Computer to Scale
Z or z
<cr> or W
U
H
R or r
10
IQ Vitals protocol (IV)
Fairbanks TeleWeigh (9600,7,ODD,2)
Scale Output
Responds (TXD)
Scale to Computer
<weight><sp><uu><sp>
<mot><sp><sp><cr><lf>
<eot>
<height><sp><hh><sp>GR<sp>
<sp><cr><lf>
Welch Allyn Spot protocol (VU)
(2400,7,EVEN,1)
Scale Output
Responds (TXD)
Scale to Computer
<mot><sp><sp>
<weight><uu><cr><lf>
Doran F0 protocol (f0)
(9600,8,NO,1)
Scale Output
Responds (TXD)
Scale to Computer
<stx><pol><weight><sp>
<uu><sp><cr><lf>
<height><sp><hh><sp>GR
<sp><cr><lf>
Description
Zero the scale (if in motion, scale will wait until
stable, then zero)
Current Display Weight. Active, Stable Hold, or
Average Hold.
( <mot> GR = stable, gr = motion )
Scale scrolls through available weight units.
Current height value in inches "in".
Scale performs a Reweigh operation (same as
pressing the Reweigh button)
Description
Current Weight.
( <mot> SD = stable, S_ = motion )
Description
Zero the scale (if in motion, scale will wait until
stable, then zero)
Current Display Weight. Active, Stable Hold, or
Average Hold.
Scale scrolls through available weight units.
Current Height value in inches or centimeters.
Scale performs a Reweigh operation (same as
pressing the Reweigh button)