- ページ 13

スイッチ Ametek esd NX30のPDF クイック・スタート・マニュアルと安全上のご注意をオンラインで閲覧またはダウンロードできます。Ametek esd NX30 17 ページ。

Ametek esd NX30 クイック・スタート・マニュアルと安全上のご注意
AMETEK CTS
5.1.2

Ports (esd NX30, NSG 438 models only)

This port is indented to allow the user to connect external signals in order to remote control the generator, as well as to con-
nect external accessories like the charge removing device. See tables and graphs below for detailed signal description and
drive circuitry information.
Connector J3: Pin Assignment
Pin
Signal Name
1
EXT_TRIGGER
2
NC
3
NC
4
EUT_FAIL
5
GND
6
NC
7
Charge remove
8
GND
9
+15V
Ext. Trigger (esd NX30 only):
The following circuit is built in behind the external trigger input connector. This function is similar to the function of the yellow
trigger knob on the handle. Trigger signal active low, i.e. to be pulled to ground.
Charge removing connection:
This function is to drive an external bleedoff switch option or an external relay.
Quick Start and Safety Guide
Description
External trigger input
-
-
EUT failure input
(reserved for future use)
Earth
-
Charge remover drive output
Earth
Voltage output (max 500 mA)
V 1.0
ESD Simulators
13 / 17