- ページ 9

ボイスレコーダー ASTRO-MED Dash 4uのPDF クイック・スタート・マニュアルをオンラインで閲覧またはダウンロードできます。ASTRO-MED Dash 4u 14 ページ。

II. Data Capture
DISCUSSION
ACTION
Enter data capture setup
Select capture of all channels
Set sample rate to 1 kHz
Set record size to 32 Ks
Set trigger window to 10%
Set capture for one record
Be sure auto-triggering is OFF
Enter trigger setup
Enable waveform triggering
Set up the logic for triggering
Pick waveform(s) for triggering
Set up the level for triggering
Run the chart
Arm for capture
Input battery voltage into Channel #1
Stop the paper
Set up one data capture for 8.19 seconds. The capture should be triggered when
the battery voltage is applied. The data capture should be a total of 8.19 seconds.
Of the total capture time, 0.819 seconds of the capture should show what the
signal looked like before the battery voltage was applied.
HOW TO
[SYS] [SETUP DATA CAPTURE]
[SET CHANNELS] Keep Channels 1 through 4.
[EVENTS] Toggle off to not capture event markers. [EXIT]
[SET SAMPLE RATE] The number of samples per second. In general, select
at least ten times the frequency of your signal or the frequency of any changes in
signal. Turn encoder wheel until sample rate is 1 kHz.
[SET RECORD SIZE] The Dash 4u will give you the amount of time that data
will be captured based on the record size and the sample rate. This is shown
under the duration field. Turn encoder wheel until record size is 32 Ks.
[SET TRIGGER Win.] Placing the trigger at any point other than the start of
the capture window will allow for pre-trigger data to be captured. Turn encoder
wheel until trigger window is 10%.
[SET OPERATION] You can capture one record or re-arm for more captures.
Turn encoder wheel for single capture.
[AUTOTRIG] If auto-triggering is ON, toggle to OFF. If toggled ON, auto-
triggering is designed to immediately capture data once the Dash 4u is armed.
[SYS] [SETUP TRIGGER]
[WAVEFORM] Waveform should be toggled to ON position.
[TOGGLE LOGIC] Insure that OR logic is chosen.
Turn encoder wheel so that only Channel #1 is selected.
[SETUP WAVEFORM] [SET CHANNEL] Turn encoder wheel to select
Channel #1. [SET TYPE / CLASS] Use wheel to select level - above for type of
trigger. [<<FIELD>>] Turn encoder wheel to 80% (8.8 volts).
[RUN / HALT]
[ARM] This key is located near the encoder wheel. The Dash 4u is now
recording pre-trigger data, waiting for the trigger.
A trigger should be produced and the capture continues to the end. While
capturing, you should have seen the waveform for Channel #1 increase from the
bottom of the channel to near the top on both the color display and the chart.
Note: you do not need to run the chart when you capture data.
[RUN / HALT]