JT2007
Pen Type Multimeter
Operating Manual
4. EF Measurement
⚠ Warning
To avoid harm to the Meter, never input higher than 300V
voltage, although it is possible to measure higher voltage.
Set the switch to V
EF and remove the test lead from
the input terminal.
Press SELECT to select EF measurement mode.
Place the positive terminals towards the object being
measured.
5. Ω
auto measurement
⚠ Warning
To avoid damages to the Meter or to the devices under lest,
disconnect circuit power and discharge all the high-voltage
capacitors before the measurement.
Set the switch to Ω
Auto measurement mode is default, in this mode, you can
measure Resistance, Diode, Continuity and Capacitance
automatically.
For better accuracy, it is better to separate the object being
measured from the circuit before the measurement.
When voltage measurement has been completed,
disconnect the testing leads from the circuit under lest.
Note:
In auto measurement mode, when input:
Resistance: <15Ω or >10MΩ
Capacitance: <400pF or >1mF
You will get an irresponsible value.
6. Resistance Measurement
⚠ Warning
To avoid damages lo the Meler or to the devices under lest,
disconnect circuit power and discharge all the high-voltage
capacitors before measuring the resistance.
Set the switch to Ω
Press SELECT to select Ω measurement mode.
Connect the test leads to the object being measured.
When resistance measurement has been completed,
disconnect the testing leads from the circuit under test, and
remove the testing leads from the input terminals of the
Meter.
7. Continuity Test
⚠ Warning
To avoid damages to the Meter or to the devices under test,
disconnect circuit power and discharge all the high-voltage
capacitors before the continuity test.
To measure resistance, do the following:
Set the rotary switch to Ω
Press SELECT to select
measurement mode.
Connect the test leads to the object being measured.
The buzzer will sound continuously if the resistance of
a circuit under test is ≤30Ω; it indicates the circuit has a
good connection.
Note:
When continuity measurement has been completed,
disconnect the testing leads from the circuit under test, and
remove the testing leads from the input terminals of the
Meler.
8. Diodes Test
⚠ Warning
To avoid damages to the Meter or to the devices under test,
disconnect the circuit power and discharge all the high-
voltage capacitors before measuring diodes.
To measure diode, do the following:
Set the rotary switch to Ω
Press SELECT to select
measurement mode.
For better accuracy, it is better to separate the object being
measured from the circuit before the measurement.
When diodes measurement has been completed,
disconnect the testing leads from the circuit under test.
9. Capacitance Measurement
⚠ Warning
To avoid damage to the Meter or to the equipment under
test, disconnect circuit power and discharge all high-voltage
capacitors before measuring capacitance. Use the DC
Voltage function to confirm that the capacitor is discharged.
To measure capacitance, connect the Meter as follows:
Set the rotary switch to Ω
Press SELECT to select
measurement mode.
For better accuracy, it is better to separate the object being
measured from the circuit before the measurement.
When diodes measurement has been completed,
disconnect the testing leads from the circuit under test.
General Specifications
Maximum voltage between red Terminals and Grounding:
300Vrms.
Maximum Display: 3,000. Updates 4 times/second
Temperature:
Operating: 3ºC~40ºC (32ºF~104ºF);
Storage: -10ºC~50ºC (14ºF~122ºF).
Relative Humidity:≤75% @ 0ºC~30ºC;
≤50% @ 31ºC~40ºC.
Altitude: Operating: 2,000m; Storage: 10,000m.
Battery Type: 3V Li-MnO2 Button cell battery.
Low Battery: Display "
".
Dimensions (HxWxL): 20.18x26.5x181.5mm.
Weight: Approx. 90g (battery included).
Accuracy Specifications
Accuracy: ±(a% reading+b digits) guaranteed for 1 year.
Operating temperature: 18ºC~28ºC.
Relative humidity: <75%.
A. AC Voltage
Range
Resolution
Accuracy
3V
0.001V
30V
0.01V
± (1%+4)
300V
0.1V
Remark: Input Impedance: ≥10MΩ
Frequency Response: 40Hz~400hZ
B. DC Voltage
Range
Resolution
Accuracy
3V
0.001V
30V
0.01V
± (1%+3)
300V
0.1V
Remark: Input Impedance: ≥10MΩ
C. EF Test
Range
Resolution
220V/50H
<10mm: buzzer beeps
10~50mm: may and may not beeps
>50mm: buzzer not sound
D. Resistance
Range
Resolution
Accuracy
300Ω
0.1Ω
3kΩ
1Ω
± (1%+3)
30kΩ
10Ω
300kΩ
100Ω
3MΩ
1kΩ
± (1.5%+5)
30MΩ
10kΩ
Remark: In the auto scan mode, the max range is 3MΩ.
E. Continuity Measurement
Range
Resolution
Remark
Open Circuit Voltage around -1.2V; buzzer
0.1Ω
beeps at resislance, ≤10Ω; buzzer not
sound when resistance is > 70Ω
F. Diodes Measurement
Range
Resolution
Overload Protection
1mV
300 Vrms
Remark: Open Circuit Voltage around 3V, displays an
approximate forward voltage drop.
G. Capacitance
Range
Resolution
3nF
0.001nF
30nF
0.01nF
300nF
0.1nF
3µF
1nF
30µF
10nF
300µF
100nF
3mF
Remark:
1. In auto scan mode, the max. range is 300μF.
2. There is a residual reading when the circuit is open. To
measure a small value of capacitance, subtract it to ensure
accuracy.
Maintenance (see figure 4)
This section provides basic maintenance information and
battery replacement instruction.
Overload Protection
300Vrms
Overload Protection
300Vrms
⚠ Warning
Do not attempt to repair or service your Meter unless you
are qualified to do so and have the relevant calibration,
performance test, and service information.
To avoid electrical shock or damage to the Meter, do not get
water inside the case.
A. General Service
Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
Clean the terminals with a cotton bar with detergent, as
dirt or moisture in the terminals can affect the readings.
Turn the Meter off when it is not in use and take out the
battery when do not use it for a long time.
Do not store the Meter in a humid, hot, explosive,
Overload Protection
inflammable place or with a strong magnetic field.
B. Replacing the Battery
⚠ Warning
300Vrms
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the
battery indicator appears.
To replace the battery:
1.
Turn the Meter off. Disconnect the testing leads from
the circuit under test, and remove the testing leads
from the input terminals of the Meter.
2.
Remove the screw from the battery compartment
and separate the battery compartment from the
case bottom.
3.
Remove the battery from the battery compartment.
4.
Replace the battery with a new 3V battery.
5.
Insert again the case bottom and the battery
compartment, and tighten the screw.
Accuracy
Overload Protection
± (3%+5)
300Vrms
± (5%+5)
Reference
Figure 4