- ページ 10
計測機器 Baumer CleverLevel PL20H/SのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Baumer CleverLevel PL20H/S 44 ページ。 Level measurement adaptive trigger – hygienic/industrial
CleverLevel
Adaptive trigger – hygienic/industrial
Configuration via remote teach
Sensors that are installed in inaccessible locations
can be easily configured via remote teach.
With the aid of qTeach, teaching of the adaptive
trigger can be in empty tank state if required
(zeroing) e.g. to suppress any interfering influence
that may be present. Restore default in the sensor
settings is also possible.
Note: Remote teach has to be activated with
FlexProgram. The configuration is only possible
during the first 5 minutes after connecting the
sensor to the power supply. Within the 5 minutes,
the switch being configured does not work.
9 The remote teach function is activated with
FlexProgrammer 9701 before the integration of
the sensor.
Short-circuit the switch output which has remote
f
teach activated with GND (0 V).
The LED rapidly blinks magenta.
Continue as described in „Configuration via
f
qTeach", step 1 and 2.
Instead of a screwdriver, use the GND.
Configuration via IO-Link master
Switch points, hysteresis, damping, output mode,
etc. can be configured through an IO-Link master.
Connect IO-Link master to sensor.
f
Connect IO-Link master to PC and set
f
parameters with the aid of BSS.
BSS is a versatile software tool for evaulation,
testing and parameterization of IO-Link devices.
Furthermore, BSS alows for connecting various
IO-Link masters, including USB-C IO-Link master.
The BSS software, the related manual and a
detailed description of IODD parameters and
process data are available for download on the
product page at www.baumer.com.
10 / 44
Level measurement
®
PL20H/S
Operating Instructions
Configuration via FlexProgrammer 9701 and
PC
Switch points, hysteresis, damping, output mode,
etc. can be configured via FlexProgram and
FlexProgrammer 9701. Furthermore, a visualization
of the measured value can be displayed by using
the online measurement in FlexProgram.
Connect the FlexProgrammer 9701 to the
f
sensor:
M12 plug version: Connect with the black head
M12 plug.
Cable version: Connect with crocodile clips.
Connect the red clip to Pin 1 (+Vs) and the
black clip to Pin 3 (GND).
Connect the FlexProgrammer 9701 to a PC with
f
FlexProgram installed.
Set parameters (refer to the HELP menu in
f
FlexProgram for more information)
www.baumer.com