1. Errors, interlocks and safeties

This section describes the error messages which can occur and some explanations on how to
deal with them.
General note: once the interlock is triggered and the error corrected, the interlock will remain
active until it the operator clicks 'Reset'. This way the user is sure to be notified and actively
allows the system to resume operation.
In case of pump diode failure, the current must be set to 0 % before restarting the diode in
order to do a slow ramp up of the power.

1.1. Magma 25

1.1.1. Safety interlocks
1.1.1.1.
The chiller for the crystal is not running or does not have high enough flow rate. Check that
the chiller is on, the water level is high enough and that the filter at the back of the chiller is
not clogged. If the water is low, refill with a mix of clean water and 10 % Optishield 1.
Filter can be cleaned by removing the tubular grid and cleaning it under tap water. If this
happens more than once or twice a month the water needs to be flushed, see corresponding
section of the start-up procedure.
1.1.1.2.
The sensor is detecting a high humidity inside the regen amplifier, which can happen if the
room humidity is too high for a long period of time or can indicate a water leakage in the
amplifier. Open the box to check for signs of water, condensation or oxidation.
1.1.1.3.
The sensor is detecting a high temperature inside the regen amplifier, which can happen if the
room temperature is too high for a long period of time or can indicate a problem with the
cooling. Check the status of crystal's chiller and the filter at the back of the chiller.

1.1. Diodes

1.1.1. Interlocks

1.1.1.1.
The chiller for the diode is not running or does not have a high enough flow rate. Check that
the chiller is on, that the water level is high enough and that the filter at the back of the chiller
is not clogged. If the water is low, refill with a mix of clean water and 10 % Optishield 1.
Filter can be cleaned by removing the tubular grid and cleaning it under tap water. If this
happens more than once or twice a month the water needs to be flushed, see corresponding
section of the start-up procedure.
2
Magma 200
Quick troubleshooting guide
Version 1.00
Flow
Humidity
Temperature
Chiller
Amplitude Laser Group
11 avenue de Canteranne, Cité de la Photonique - Bâtiment MEROP A 33600 Pessac - France
Headquarters
/
Capital social: 317 100 €
/
/
Siret 434 637 005 - APE 6420Z
Tel: +33 (0)5 56 46 40 60
/
Fax: +33 (0)5 56 46 06 94
Headquarters
/
11 avenue de Canteranne, Cité de la Photonique - Bâtiment MEROPA 33600 Pessac - France
Capital social : 320 925 €
Amplitude Laser Group
/
Tel: +33 (0)5 56 46 40 60
/
N°TVA Intracom: FR 70 434 637 005
/
/ [email protected] / www.amplitude-laser.com
/
SIRET: 434 637 005 - APE 6420Z
/
N°TVA Intracom: FR 70 434 637 005
Fax: +33 (0)5 56 46 06 94
/
/ www.amplitude-laser.com