Pressing P the normal visualization is restored.
Free delivery: it is possible to exclude the liter-counter (for example in case of malfunction of the
part) by means of the special program P99:
With the START and STOP keys the electrovalve opens and closes.

ERROR MESSAGES AND CORRECTIVE ACTIONS

3.5
All error messages are accompanied by flashing wordings and an intermittent acoustic signalling. The
latter one can be stopped with the STOP key. Using the START key the error message is cancelled and
the delivery starts again. On the contrary, the double STOP command cancels the delivery. The possible
error messages and relevant wordings are listed below.
3.5.1
Quantity counting error
After pressing START, after 12" this wording appears:
Possible causes:
A) If no water comes out of the service
The cocks on the inlets of warm water and/or cold water are not open. Check that both are open and
that the connection to the pipes is correct.
The inlet filters are obstructed. Clean them.
The electrovalve dos not open. Contact the service.