- ページ 8
計測機器 Doza DKG-05DのPDF 操作マニュアルをオンラインで閲覧またはダウンロードできます。Doza DKG-05D 17 ページ。 Personal gamma radiation dosimeter
-
if measured rate of
duration would be switched on to signalize rate of
such case maximal possible value of 42.9 Sv/h would be indicated;
-
when the accumulator battery voltage is below 3,52 V, the blinking battery symbol
would be indicated to prevent that the dosimeter will be automatically switched off
in few hours because of the complete power discharge; in such case 2 second sound
signal would be given each 15 minutes. In case of 3,3 V battery voltage, the switching
off signal will be given and the dosimeter will be automatically switched off with the
preservation of accumulated information in the memory. In case of dosimeter operation
under negative outside temperature around minus 20 °С, the battery discharge symbol
can be also indicated for almost charged battery. In such case the automated switching
off will occur after 24 hours of continuous operation at least.
Rate of
10
p
automatically increased for the decreasing dose rate. Each sequential calculation of rate of
value is accompanied by sound signal of 60 ms duration.
Dosimeter power supply is provided by three accumulator batteries of no less than
280 mA/h capacity each with the total supply voltage of less than 4,2 V. Continuous operation
without battery charge is 100 hors for routine operation of the sound signal: less than 1 sound
signal of measurement per minute. Battery charge is provided by КЗУ-27 charger included in the
shipment set.
For dosimeter operation in the dosimetry monitoring system together with PC, the data
transmission is arranged via УС-05С.
1.5 Marking and sealing
1.5.1 Following marks are placed on the casings of dosimeter components:
-
trade mark or name of the manufacturer;
-
conditional sign of item: item type and unit;
-
sequential number of item according to the manufacturer;
-
year of production;
-
approval sign of measuring instrument pattern (only for the dosimeter);
-
power, voltage/current and electric power frequency;
-
protection degree on ГОСТ 14254-96.
1.5.2 The place and technology of marking and font size agree to requirements specified
in design documentation.
1.5.3 The dosimeter, УС-05С and КЗУ-27 must be sealed according to design
documentation.
1.6 Packing
1.6.1 Dosimeter packing should be done according to the requirements of category КУ-3
on ГОСТ 23170-78.
1.6.2 Inner package of the dosimeter, УС-05С and КЗУ-27 should correspond to the
requirements of ГОСТ 9.014-78 for group III of ВЗ-10 protection type, ВУ-5 package variant.
Protection term without re-conservation is 3 years.
1.6.3 Packing should be carried out indoors in the ventilated rooms under the air
temperature of +15 to +40 °С, relative air humidity of up to 80 % at +20 °С, and air
concentration of corrosion agents according to type I on ГОСТ 15150-69.
10
value is above 42.9 Sv/h, the continuous alarm of 6 seconds
p
measurement time is 1 to 255 s. The time of indication renewal is
ФВКМ.412113.005РЭ
10
indication limit excess; in
p
10
p
8