- ページ 12

計測機器 Doza MKS-15D SnegirのPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Doza MKS-15D Snegir 17 ページ。 Dose-rate meter / radiometer

Doza MKS-15D Snegir ユーザーマニュアル

2.4 Adjustment

2.4.1 Change of battery type
To change the battery type, turn on the dosimeter while holding the
message "SELECT THE TYPE OF BATTERY" and currently used battery type will be shown on the
display. Use
and "MODE" buttons to select the desired type; confirm selection by pressing
"START" button in this mode; after that the dosimeter will switch over to the working mode. Use
button to exit the selection mode. At that, the dosimeter will switch over to the working mode and the
previous type of batteries will remain unchanged.
2.4.2 Change of the confidence probability value (predetermined reliability)
To change the value of confidence probability, switch over to the engineering mode. For that,
turn the dosimeter on while holding the
A request for entering the access code will appear on the display. The code is as follows: "START"->
"MODE"-> "MODE"-> "START"-> "START". Use
value. Use "START" button to confirm or
off. Turn the dosimeter on, make sure that required value is set (see symbol in the message line).
2.4.3 Correction of adjusting coefficients
Adjusting coefficients are set by manufacturer and can be changed only during repair or
calibration in the engineering mode.
2.4.3.1 Alternative assignment of controls
Alternative assignment of controls in the engineering mode is presented in the Table 2.4.
Table 2.4
Function (name) of buttons
in working mode
"MODE"
"START"
2.4.3.2 For review of current working coefficients turn on the dosimeter while pressing the
button and, holding it pressed, read data on the display (see Table 2.5).
Table 2.5
Indicated data
ALKALINE
Battery type used for calculation of residual capacity
ADC correction coefficient
K
104
v
Correction coefficient for the detector's background
K
0.00e+00
f
K
2.20e-07
Conversion coefficient
x
"Dead time"
K
2.20e-05
m
K
1
1.45e+01
Transition coefficient to beta radiation flux density
К
2
0.00e+00
Conversion coefficient 2
К
3
0.00e+00
Conversion coefficient 3
2.4.3.3 Correction of workout coefficients
For correction of workout coefficients switch to the engineering mode; for that, turn the
dosimeter on holding the following buttons pressed simultaneously:
, "MODE" and "START" buttons pressed simultaneously.
and "MODE" buttons to select the required
button for cancel the entered value, turn the dosimeter
Function (name) of buttons
in engineering mode
Meaning
FVKM.412152.005RE
12
"↑"
"↓"
"ESC"
"ENTER"
, "MODE" and "START".
button pressed; a