3. Maintenance
3.1 Meter maintenance
It is forbidden to separate the housing of the instrument.
In addition to the occasional need to use a damp cloth to wipe, the instrument does
not need to do other maintenance. The shell is made of plastic and is subject to
erosion by some organic solvents such as toluene, xylene and butanone. If this is
the case, immediately wipe the spill onto the shell of such solvents.
3.2 Electrode maintenance
Ensure that the electrode is always stored in the appropriate storage solution.
In order to obtain maximum accuracy, any filled or coagulated in the electrode
outside the filling liquid are used to remove the distilled water in time. Always
store the electrode according to the manufacturer's regulations and do not dry
it. If the slope of the electrode drops rapidly, or if the corresponding speed is
slow, the following steps can be used. Depending on the sample, try the
following:
1. For grease, remove the dirt from the surface of the electrode with raw
cotton dipped in acetone or soapy water.
2. If the electrode film is dry, immerse the electrode tip in 0.1M HCl solution
and allow to stand overnight.
3. If protein is accumulated in the membrane, immerse the electrode in the
HCI/pepsin solution to remove the deposit.
- 11 -