Troubleshooting the Current Tracer

To test the operation of the transmitter, connect it to a known power source from 6 to 300 V ac/
dc. The LED should begin to blink at a steady rate of approximately 71 times per minute. If it
does not blink, checks to ensure that there is proper voltage on the line and replace the fuse,
if necessary. If the LED still does not blink after checking the voltage and the fuse, send the
transmitter In for repair.
Select "Breakers-5" and hold the receiver's tip to the front and center of the transmitter. Push
the "ON" button and wait 2 seconds for the probe to calibrate. After calibration, 10 LEDs should
flash from left to right in sync with the LED on the transmitter. If it does not, the battery may be
low. After replacing the battery and the LEDs still do not flash, send the probe in for repair.

SPECIFICATIoNS

operating frequency:
operating temperature:
Storage temperature:
T300 Transmitter fuse:
Case operating voltage:
operating frequency:
Line frequency range:
Maximum load:
duty cycle:
P26 Receiver
Power:
battery life:
Case:
50/60 Hz Rejection:
Accessories and replacement parts
Pigtail connector:
Power Cord (T300):
Adapter:
Earphone, 3.5mm:
Fuse, ¼ amp, 250V, Fast Acting: 3AG-312.250
Carrying case:
1.800.561.8187
6.25 kHz
0°F to 120°F (-18°C to 49°C)
-40°F to 150°F (-40°C to 66°C)
¼ amp, 250V, fast acting (5x20mm)
Flame resistant ABS plastic 6-300 V ac/dc
6.25 kHz
dc to 1 kHz
200 mA peak at 110 V
.164 second every .82 second
9 V Alkaline battery IEC #6LR61
100 hours (typical)
Flame resistant ABS plastic
-156 dB to 20K amps @ 50/60 Hz
PT26
PC-326
AD-1
EP1001
CC-326