- ページ 12
マルチメーター Amprobe 34XR-AのPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Amprobe 34XR-A 19 ページ。 Professional digital multimeter
RESISTANCE
DUTY CYCLE
Ω
Ω
Ω
Ω
Ω
Ranges: 400
, 4k
, 40k
, 400k
, 4M
,
Ranges: 0 to 90 %
Ω
40M
Resolution: 0.1 %
Ω
Resolution: 100 m
Pulse widh: >10 µs
Accuracy: ±(1.0 % rdg + 4 dgts) on
Frequency range: 40 Hz to 20 kHz
Ω
Ω
400
to 4M
ranges
Accuracy: (5V logic ) ±(2.0% rdg + 5 dgts)
Ω
±(2.0 % rdg + 5 dgts) on 40M
range
Overload protection: 1000 V dc or
Open circuit volts: -0.45 V dc typical,
750 V ac rms
Ω
(-1.2 V dc on 400
range)
CONTINUITY
Overload protection: 1000 V dc or
Audible indication: < 35
750 V ac rms
Response time: 100 ms
CAPACITANCE
Overload protection:1000 V dc or
Ranges: 4µF, 40µF, 400µF, 4000µF
750 V ac rms
Resolution: 1 nF
DIODE TEST
Accuracy:
Test current: approximately 1.2 mA
±(5.0 % rdg + 10 dgts) on 4µF range
Accuracy: ±(1.5 % rdg + 3 dgts)
±(5.0 % rdg + 5 dgts) on 40µF to
Resolution: 1 mV
400µF ranges
Open circuit volts:
±(5.0 % rdg + 15 dgts) on 4000µF range
Overload protection: 1000 V dc or
Test voltange: < 3.0 V
750 V ac rms
Test Frequency: 25Hz
ADDITIONAL FEATURES
Input protection: 0.315A/1000V fast blow
µA mA
ceramic fuse 6.3×32mm on µA/mA input
to warn test leads are connected to
TEMPERATURE
measure current while Function/Range
Ranges: -20 °C to 1000 °C, -4 °F to 1832 °F
Switch is not set to a measure current.
Resolution: 1 °C, 1 °F
MIN MAX: Displays the minimum or
maximum value detected while making a
Accuracy:
measurement.
±(2.0 % rdg + 4 °C) -20 °C to 10 °C
HOLD: Holds the latest reading on the
±(1.0 % rdg + 3 °C) 10 °C to 200 °C
display.
±(3.0 % rdg + 2 °C) 200 °C to 1000 °C
RANGE:
±(2.0 % rdg + 8 °F) -4 °F to 50 °F
Backlight:
±(1.0 % rdg + 6 °F) 50 °F to 400 °F
approximately 60 seconds
±(3.0 % rdg + 4 °F) 400 °F to 1832 °F
Auto Power off: 30 minutes, typical
Overload protection: 1000 V dc or
REPLACEMENT PARTS
750 V ac rms
TL36
FREQUENCY
FP300 mA fuse - Fuse Pack .315A/1000V
Ranges: 4k, 40k, 400k, 4M, 40MHz
Resolution: 1 Hz
FP100 10A fuse - Fuse Pack 10A/1000V
Accuracy: ±(0.1 % rdg + 3 dgts)
Sensitivity:
XR-TA Input Adapter for K-type
10 Hz to 4 MHz: >1.5 V ac rms;
4 MHz to 40 MHz: >2 V ac rms, <5 V ac rms
TP255 K type thermocouple
Minimum pulse width: > 25 ns
XR-H2 Magne-Grip
Duty cycle limits: > 30 % and < 70 %
Overload protection: 1000 V dc or
750 V ac rms
11
Ω
3.0 V dc typical
, 10A Test Lead Connection: Beeps
Manual range mode.
Backlight auto-off
Test Lead Set with Alligator clips
(4 each)
(2 each)
thermocouple
®
Holster, clip, magnet,
and strap