- ページ 2

温度コントローラー Carbolite 201のPDF 操作説明をオンラインで閲覧またはダウンロードできます。Carbolite 201 4 ページ。
Carbolite 201 にも: 取扱説明書 (8 ページ)

1. O
& K
UTLINE
When switched on, the controller lights up, goes through a short test routine, and then displays
the measured temperature and starts to control. The output light glows or flashes as heating
occurs.
The Page  key allows access to parameter lists within the controller; most lists and parameters
are hidden and cannot be accessed by the operator because they contain factory-set parameters
which should not be changed. A single press of the page key displays the temperature units,
normally set to °C; further presses reveal the lists indicated in the Navigation Diagram below.
The Scroll  key allows access to the parameters within a list. A single press displays the
temperature units; further presses reveal the parameters in the current list indicated in the
Navigation Diagram. Some parameters are display-only; others may be altered by the operator.
To return to the Home list at any time, press Page and Scroll together, or wait for 45 seconds.
The Down  and Up  keys are used to alter the setpoint or other parameter values.
2. 201 N
AVIGATION
Home
List
20.0
°C/F/k
OP
wSP
SPrr
dwEll
MC01A 1.01
EYS
D
IAGRAM
Output
List
°C/F/k
oP
Measured
OP.Hi
temperature; use
Down/Up to
access setpoint
OP is a read-only value showing the current
output power level
wSP (Working Setpoint) is only present if
the ramp rate SPrr is active
201T only
only present when
access to the Power
Limit may be required
201: PID controller
with ramp to
setpoint feature
201T: 201 with
Process Timer
option fitted
Access
List
ACCS
codE
Used in factory
configuration; not
accessible to the operator
home list