(contrast or definition
isn't enough)
One side of image is
dark or image is moving
as focusing
Objective touches
specimen as changing
low times objective to
high times objective
Image observed by two
eyes aren't in
superposition entirely.
It is easy for eyes to be
tired during observing
12
Surface of objective lens is
dirty (especially it is easy for
the front lens of 40X
objective to dip in immersion
oil)
Immersion oil isn't used for
100X objective (oil)
Immersion oil doesn't meet
the requirement
There is bubble in immersion
oil
Size of iris aperture isn't
proper
Position of condenser is too
low
Objective isn't in correct
position
Specimen isn't placed
correctly
Cover glass of specimen isn't
in up direction
Cover glass doesn't meet the
requirement
Interpupilary distance isn't
adjusted correctly
Diopter isn't adjusted
correctly
Clean it
Use immersion oil
Use immersion oil
supplied by us
Clear the bubble way
Adjust the size of iris
aperture
Readjust the position of
condenser
Make the objective in
correct position
Place specimen levelly
on stage and clip it with
clamp
Place specimen correctly
Use required thickness
cover glass (0.17mm)
Adjust interpupilary
distance according to two
eyes
Readjust diopter