Table 206 – Recommended separation distances between portable and mobile
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING (see 6.8.3.201 b))
portable and mobile RF communications equipment and the MACH 120 / 120 F
The MACH 120 / 120 F is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the MACH 120 / 120 F can help prevent electromagnetic interference by maintaining a minimum distance be-
tween portable and mobile RF communications equipment (transmitters) and the MACH 120 / 120 F as recommended below, according
to the maximum output power of the communications equipment
Rated maximum output
of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
59080001
Ceiling, Wall, Stand model
RF communications equipment and the EQUIPMENT or SYSTEM -
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
=
d
, 1
17
P
0,12
0,37
1,17
3,69
11,67
Edition 08
m
80 MHz to 800 MHz
=
d
, 1
17
P
0,12
0,37
1,17
3,69
11,67
06.02.2010 / Bak Page 19/19
Dr. Mach
Lamps and Engineering
800 MHz to 2,5 GHz
=
d
, 2
34
P
0,23
0,74
2,33
7,38
23,33