照明器具 Dr. Mach Mach M2 SeriesのPDF 取付方法をオンラインで閲覧またはダウンロードできます。Dr. Mach Mach M2 Series 19 ページ。 Stand lamp

Stand for Mach M2
The stand foot must not be connected to the mains, until it has been fully assembled
and installed. Connect the stand only to a correctly installed socket with earthing con-
tact. The socket used must be installed according to the requirements of IEC and VDE
0107.
This device is not suitable for use in hazardous locations.
The devices are not suitable for use in combustible blends of anaesthesia agents with
oxygen or nitrous oxide.
Repairs to the stand and special installation work on the sliding contacts should only
be carried out by ourselves or a company expressly authorised by ourselves.
The manufacturer is only responsible for operational safety of the stand foot when re-
pairs and modifications are carried out by his own staff or by persons who guarantee
compliance with the safety regulations.
The manufacturer is not responsible for damage to persons or property caused by in-
correct or improper use, or when used for incorrect purposes.
General information
These installation instructions must be kept at hand together with the relevant
operating instructions for consultation at any time.
The stand is designed for carrying and positioning of medical examination lamps up to a max.
weight of 4,5 - 5kg and to supply the lamps with electricity.
The stand is available:
with built-in transformer 22,8V sec.; 50/60Hz
The output voltage of the transformer is not adjustable.
The device is designed for continuous operation.
Make sure, that the maximum load does not exceed 5kg.
Please read these mounting instructions and directions for use carefully before working with the
device. In this way you will profit from all advantages of the device and save yourself and other
persons from harm.
59860007