- ページ 3
ラボラトリー機器 Drawell DW-FD SereriesのPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Drawell DW-FD Sereries 17 ページ。
Precautions on Safe Operation
Precaution:
♦
For your safety and to operate the instrument properly, please read this operation instruction
carefully before using and operate the following requirements in this manual.Where you violate
the instruction of manufacturer in this manual, it is possible to damage the protection of this
instrument.
♦
This instrument shall not be used for purposes other than sterilization
not be used for sterilizing the flammable, explosive and readily oxidizable products or strong acid,
alkali and saline solutions to avoid explosionand corruption of sterilizer chamber or pipelines.
♦
The power supply of sterilizer should be properly connected according to nameplate on the
instrument. In case of excessive voltage fluctuation, the regulated power supply shall be used to
ensure optimum performance of the instrument. IF you are using other types of voltage, a
transformer must be used to prevent the instrument from being damaged.
♦
It is necessary to ground the instrument properly, and never connect the ground wire to
plastic pipeline, gas pipe, telephone ground wire and lightning rod.
♦
Never block the steam exhaust hole on the safety valve with articles, so that the safety valve
may work properly under abnormal conditions.
♦
Always ensure that the reading of pressure gauge is "OMpa" before opening the chamber
cover. Never attempt to open the chamber cover and drainage valve, or draw out the steam
trapping bottle when the pressure inside the sterilizer chamber is over "OMpa" to avoid injury
from ejected pressurized steam.
♦
Avoid dripping water to the control circuit when adding distilled water to the sterilizer
chamber to against electric shock or other faults.
♦
When using waste processing bags or other type of bags, please put the bags into the
stainless steel basket before put into the sterilizer chamber to avoid affecting the precision of
temperature.
♦
If the handle is turned by mistake during sterilization, it will not continue turning as it has
been locked by the interlocking device, and there will be a little steam exhausted to release. In this
case, press the"STOP" button to stop the instrument immediately. When opening the chamber
cover, turn the handle slightly toward the opposition direction that it is wrongly operated, and
then turn the handle counter-clockwise until the cover is released.
♦
Observe the temperature variation inside the sterilizer chamber. The temperature is relatively
high when operation has been finished, keep the face and hands away from the sterilizer chamber
when opening the cover to avoid injury from ejected pressurized steam. Gloves must be used
when taking articles out of the sterilizer chamber.
♦
Liquid does take some time to get cooling, verify whether the temperature is low enough or
not, when taking the liquid after sterilization out of the chamber to avoid scalding.
Drawell Scientific
drying, and it shall also
,