- ページ 8
ラボラトリー機器 Drawell DW-FD SereriesのPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Drawell DW-FD Sereries 17 ページ。
Chapter 3 Installation of Instrument
Placement of Instrument
1.Place the instrument on the level ground with the brakes of four casters pressed down
before installation.Do not place any objects within 10cm around for ventilation.
2.Connect the drainage pipe between machine and wastewater
receiving bottle or pipe.
3. Take out the water plate,remove the packaging, place it in the bottom of the chamber
in a horizontal position.
Connection
instrument must be grounded reliably, if the power socket does not
1. The instrument must be grounded reliably, if the power socket does not have the
ground terminal, it is required to connect the instrument with independent ground
wire before powered on.
2.Power supply: single-phase AC 220V
3. Requirement for current intensity: FD36A/FD36R 16A
Drawell Scientific
土
10%, 50Hz/60Hz
FD50A/FD50R
FD65A/FD65R
FD80A/FD80R
FD100A/FD100R 32A
16A
20A
32A