- ページ 10
ラボラトリー機器 Drawell DW-TDZ5-WSのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Drawell DW-TDZ5-WS 20 ページ。 Benchtop low-speed centrifuge
and put it on the rotor base vertically and placidly while the rotor bottom's
location cone aims at the location cone of drive shaft; then put the flat gasket
on the lead screw and screw down the cap nut. At last, hold up the rotor
steadily (not shacking) by one hand, and put the solid wrench on the cap nut
by another hand. In the mean while turn the wrench for 3 to 6 rounds in
clockwise direction until the rotor connects the drive shaft closely.
Rotor disassembly: (After finishing adjust or centrifugal)
First clasp the spoke of rotor with one hand, and put the solid wrench on the
cap nut by another hand. In the mean while turn the wrench for 3 to 6 round in
counter clock wise direction to loose the cap nut until the rotor can be
separated from driving shaft. If the rotor is not taken out from the drive shaft for
a lone time and blocked, you can beat the surrounding spoke of rotor several
times lightly with wooden or rubber hammer. Then clasp the symmetric spoke
of rotor and lift slightly by both hands, and put the rotor on the flat table and
ground covered with soft pad.
Warning: All actions on the rotor must be soft for avoiding the damage of
driving shafting.
6. operation
6.1 Safety Notice
When the centrifuge is under high-speed operation, a small mistake of
8