above misoperation should be borne by user.
* please wipe up the rotor with neutral cleaning mixture after taking out for
avoiding corrosion and then stock it at dry, ventilated place. Central hole in
rotor should be smeared some lubricating grease for protection.
* All action on rotor should be soft, and rotor should be taken out vertically for
protecting the shaft.
* Please wipe up the water in centrifugal chamber if it is leaving unused. Some
lubricating grease should be smeared on drive shaft for protection.
* The interval of between two operations is no less than 5 minutes for avoiding
damage of compressor.
* If rotor damage occurring, allowable space to move for centrifuge is 30cm. *
Anything should not be put around centrifuge within 30cm, especially for
flammable and explosive matter.
* It is not allowed to lean on the centrifuge when it is running. Non-staff are not
allowed to stay in the operation area.
* When unpacking rotor and replacing rotor accessory and sample under
low-temperature, the operator must adapt protection actions for avoiding tissue
frostbiting.
* Centrifuge must be stopped when ant abnormal phenomena occurring or
error code displaying, then you should inform our company or service agent
soon.
Attention for safety in chemistry and biology
The routine manipulation may include all kinds of liquid and test sample which
may be the disease-causing and poisonous matters. The centrifuge can not be
manipulated with the matters unless you have taken the protection measures.
* Please pay attention to the liquid explanation in container before operation.
* Be careful of the infectious liquid. If not understanding the provided sample.
Please at least make sure that they aren't microbe. You must make further
aerosol protection (the suspended speck aren't in the air, such as smoke and
fog). For the easier disease-causing hepatitis virus (B or C), HIV, atypical
myco-bacteria and the fungus of physical structure.
* The infectious samples must be manipulated according to the manipulating
3