8.3 Digital inputs

The free voltage digital input is programmable in different configurations by the "iF" parameter.
Door switch
It signals the door status and the corresponding relay output status through the "dC" parameter: no
=normal (any change); Fn = Fan OFF; CP = Compressor OFF; FC = Compressor and fan OFF.
Since the door is opened, after the delay time set through parameter "di" , the door alarm is enabled,the
display shows the message "dA" and the regulation restarts if rd = y. The alarm stops as soon as the
external digital input is disabled again. With the door open, the high and low temperature alarms are
disabled.
External alarm(iF=EA)
As soon as the digital input is activated the unit will wait for "di" time delay before signalling the "EA"
alarm message. The outputs status don't change. The alarm stops just after the digital input is de-activated.
Serious alarm (iF=bA)
When the digital input is activated, the unit will wait for "di" delay before signalling the "CA" alarm
message. The relay outputs are switched OFF. The alarm will stop as soon as the digital input is de-
activated.
Start defrost(iF=dF)
It starts a defrost if there are the right conditions. After the defrost is finished, the normal regulation will restart
only if the digital input is disabled otherwise the instrument will wait until the "dd" safety time is expired.
Inversion of the kind of action: heating-cooling (iF=Hc)
This function allows to invert the regulation of the controller: from cooling to heating and viceversa.

8.4 Alarm signaling

Mess.
"P1"
Room probe failure
"P2"
Evaporator probe failure
"HA"
Maximum temperature alarm
"LA"
Minimum temperature alarm
"EA"
External alarm
"CA"
Serious external alarm
"dA"
Cause
Door Open
Outputs
Compressor output according to "Cy" e "Cn"
Defrost end is timed
Outputs unchanged
Outputs unchanged
Outputs unchanged
All outputs OFF.
Compressor and fans restarts
10