- ページ 17

医療機器 BD 9003TIG03-GのPDF クイック・リファレンス・マニュアルをオンラインで閲覧またはダウンロードできます。BD 9003TIG03-G 18 ページ。 Volumetric pump

BD 9003TIG03-G クイック・リファレンス・マニュアル
Display
DOSE WOULD EXCEED
Infusion rate entered exceeds a Guardrails soft
limit.
DOSE UNDER
Infusion rate entered is under a Guardrails soft
limit.
DOSE NOT PERMITTED
Dose rate entered exceeds the hard maximum
limit.
RATE NOT PERMITTED
Infusion rate entered exceeds the hard
maximum limit.
CONCENTRATION NOT PERMITTED
Concentration entered exceeds the hard
maximum limit, or is under the hard minimum
limit.
BOLUS DOSE NOT PERMITTED
Bolus dose rate entered exceeds the hard
maximum limit.
WEIGHT ABOVE LIMIT
Patient weight entered exceeds a Guardrails
soft limit.
WEIGHT BELOW LIMIT
Patient weight entered is under a Guardrails
soft limit.
BOLUS DOSE WOULD EXCEED
Bolus dose rate entered exceeds a Guardrails
Soft Limit.
BOLUS DOSE UNDER
Bolus dose rate entered is under a Guardrails
soft limit.
0000CF04424.indd 17
Action
• Check the infusion setting.
• To confirm OVERRIDE LIMIT?
press YES.
• To deny OVERRIDE LIMIT?
press NO.
• Check the infusion setting and
adjust to appropriate required
value.
• Check the weight setting.
• To confirm OVERRIDE LIMIT?
press YES.
• To deny OVERRIDE LIMIT?
press NO.
• Check the bolus setting.
• To confirm OVERRIDE LIMIT?
press YES.
• To deny OVERRIDE LIMIT?
press NO.
0000CF04424 Iss. 1 17/18
17/09/2019 10:45:06