Operating Manual
Digital Field Display PA 440
www.bdsensors.cz
BA_PA440_E_SRO
Headquarters Eastern Europe
Headquarters Western Europe
BD SENSORS s.r.o.
BD SENSORS GmbH
Hradištská 817
BD-Sensors-Str. 1
CZ - 687 08 Buchlovice
D - 95199 Thierstein
Czech Republic
Germany
Tel.: +42 (0) 572-4110 11
Tel.: +49 (0) 9235-9811-0
Fax: +42 (0) 572-4114 97
Fax: +49 (0) 9235-9811-11
Russia
China
BD SENSORS RUS
BD SENSORS China Co, Ltd.
39a, Varshavskoe shosse
Room B, 2nd Floor, Building 10,
RU - Moscow 117105
No. 1188 Lianhang Rd.
Russia
201112 Shanghai,
Tel.: +7 (0) 95-380 1683
China
Fax: +7 (0) 95-380 1681
Tel.: +86 (0) 21-51600 190
Fax: +86 (0) 21-33600 613
further agencies in:
EUROPE
• Belgium
• Poland
• Denmark
• Romania
• France
• Sweden
• Greece
• Switzerland
• Italy
• Slovakia
• Lithuania
• Spain
• Luxemburg
• Turkey
• Netherlands
• UK
• Norway
• Ukraine
The addresses of our distribution partners are listed on our
homepage www.bdsensors.com. It is possible to download
data sheets, operating manuals, ordering codes and certifi-
cates, as well.
1. General information
1.1 Information on the operating manual
This operating manual contains important information on
proper usage of the device. Read this operating manual
carefully before installing and starting up the pressure
measuring device.
Adhere to the safety notes and operating instructions which
are given in the operating manual. Additionally applicable
regulations regarding occupational safety, accident preven-
tion as well as national installation standards and engineer-
ing rules must be complied with!
This operating manual is part of the device, must be kept
nearest its location, always accessible to all employees.
This operating manual is copyrighted. The contents of this
operating manual reflect the version available at the time of
printing. It has been issued to our best knowledge. However,
errors may have occurred. BD SENSORS is not liable for any
incorrect statements and their effects.
– Technical modifications reserved –
1.2 Symbols used
DANGER! – dangerous situation, which may result in
death or serious injuries
WARNING! – potentially dangerous situation, which
may result in death or serious injuries
CAUTION! – potentially dangerous situation, which may
result in minor injuries
!
CAUTION! – potentially dangerous situation, which may
result in physical damage
NOTE – tips and information to ensure a failure-free
operation
1.3 Target group
WARNING! To avoid operator hazards and damages of
the device, the following instructions have to be worked
out by qualified technical personnel.
1.4 Limitation of liability
By non-observance of the operating manual, inappropriate
use, modification or damage, no liability is assumed and
warranty claims will be excluded.
1.5 Intended use
- The field display PA 440 is supplied by the analogue
current loop and shows the measured value on the dis-
play. For monitoring the limit values optionally up to two
PNP open collector-contacts are available. As standard
the PA 440 is equipped with a LC-display, optionally a
LED-display is deliverable.
- The configuration is menu-driven via two push buttons
located in the front. Following parameters could be con-
figured: decimal point, zero point, end point, switch-on
and switch-off points, etc. Those parameters are being
stored in an EEPROM and, thus, are being kept also in
case of power breakdown. Limit exceeding in both direc-
tions can be displayed as a message. Furthermore an
access protection is provided.
- It is the operator's responsibility to check and verify the
suitability of the device for the intended application. If
any doubts remain, please contact our sales department
in order to ensure proper usage. BD SENSORS is not
liable for any incorrect selections and their effects!
- The technical data listed in the current data sheet are
engaging and must be complied with. If the data sheet is
not available, please order or download it from our
homepage. (http://www.bdsensors.com/products/down-
load/datasheets)
WARNING! Danger through improper usage!
1.6 Package contents
Please verify that all listed parts are undamaged included in
the delivery and check for consistency specified in your
ASIA
order:
• Iran
- field display PA 440
• Israel
• Kazakhstan
- sheet of unit labels
• Taiwan
- mounting instructions
• Thailand
2. Product identification
AUSTRALIA
The device can be identified by its manufacturing label. It
provides the most important data. By the ordering code the
product can be clearly identified. The programme version of
the firmware, (e. g. P06) will appear for about 1 second in
the display after starting up the device. Please hold it ready
for inquiry calls.
nominal
pressure
range
signal
Fig. 1 manufacturing label- example
!
The manufacturing label must not be removed from the
device!
3. Mechanical installation
3.1 Mounting and safety instructions
WARNING! Install the device only when currentless!
WARNING! This device may only be installed by
qualified technical personnel who has read and under-
stood the operating manual!
!
Handle
measuring device with care, both in packed and
unpacked condition!
!
There are no modifications/changes to be made on the
device.
!
Do not throw the package/device!
type
serial
designation
number
ordering code
this
high-sensitive
electronic
precision
!
Remove packaging only directly before starting up the
device to avoid any damage!
!
Do not use any force when installing the device to
prevent damage of the device and the transmitter!
3.2 General installation steps
- Carefully remove the pressure measuring device from
the package and dispose of the package properly.
- Next the field display has to be mount stationary on a
suitable fixing location via two appropriate fastening
screws.
3.3 Lead in the transmitter cable
-
Lead in the connecting cable of the transmitter through
the cable gland on the left side. The cable length inside
the terminal box has to be long enough for connecting
the cords with the terminal clamps on the left
(SENSOR).
-
Then tighten the cable gland by hand. Take care that
the cable is strain-relieved.
-
Pay attention that the PTFE-filter on the gauge refer-
ence of BD SENSORS-transmitters may not be dam-
aged or removed.
3.4 Lead in the supply line
-
Lead the supply line through the cable gland on the
right side. The cable length inside the terminal box has
to be long enough for connecting the cords with the
terminal clamps on the right (SUPPLY).
-
Then tighten the cable gland by hand. Take care that
the cable is strain-relieved.
4. Electrical Installation
WARNING! Install the device only when currentless!
Open the top cover; establish the electrical connection of
the device according to the following table and the wiring
diagram. Screw the top cover onto the box again.
Pin configuration:
Electrical
De-
connection
Terminal
sign-
(cable colours of
block
ation
BD SENSORS
transmitters)
Potential reference
GND SENSOR
clamp
(yellow/green )
VS- SENSOR
Supply - (brown)
VS+ SENSOR
Supply + (white)
SP2
SP
Contact 1
SP1
SP
Contact 2
VS+
SUPPLY
Supply +
VS-
SUPPLY
Supply -
Potential
GND SUPPLY
reference clamp
For the electrical connection a shielded and twisted
multicore cable is recommended.
The ground wires of all components have to be
connected when installing!
Supply:
The supply created by the electronics of the device is
approx. 6.5 V
. Please take this into consideration when
DC
planning your power supply. The tolerances for the power
supply can be calculated as follows:
V
= V
+ 6.5V
Minimum supply:
Smin
minTR
V
= V
+ 6.5V
Maximum supply:
Smax
maxTR
V
= minimum supply of the used 2-wire transmitter
min TR
V
= maximum supply of the used 2-wire transmitter
maxTR
5. Initial start-up
WARNING! Before the initial start-up, the user has to
check that the device has been properly installed and
make sure that it does not have any visible defects.
WARNING! The device must only be started up by
qualified technical personnel that has read and under-
stand the operating manual!
WARNING! The device must only be used within the
technical specifications (compare the data in the data
sheet)!
to connected with
cable shield of
transmitter
neg. connecting
cable of transmitter
pos. connecting
cable of transmitter
connecting cable
of transmitter
for contact 1
connecting cable
of transmitter
for contact 2
pos. connecting
cable for pressure
signal
neg. connecting
cable for pressure
signal
cable shield of
supply line