6. Wipe X Galvo Mirror
Prepare a new swab. It is
important
to use a fresh swab so as not to spread
any particles or residues that had been caught by the previous swab.
Carefully position the galvo block as shown. Use the swab to gently rotate the
mirror so that the surface is more easily accessible. Gently wipe the mirror
from the base of the mirror to the end of the mirror. Rotate the swab and
wipe again.
7. Inspect the Mirrors
.
Once the mirrors are clean, the galvo block can be reinstalled into the Form
1+. Use a bright light or flashlight to illuminate the mirrors.