4.
Tighten the four socket head cap screws [I] to
secure the sample cooling assembly and the cable
clip.
5.
Lift up the movable plate that secures the sample
probe, and lower the sample well mounting block.
6.
Remove the sample tube from the sample cooling
assembly.
7.
Secure the thermo‑electric wire leads with the cable
clip [K].
8.
Secure the block probe cable, the thermo‑electric
cable, and the solenoid cable wire leads using the
two cable clamps [J].
Make sure that there is a service loop [O] between
the two cable clamps.
9.
Reconnect the sample well drain tube [G] from the
fitting located on the drain tube adapter [H].
10.
Secure the cooling fan, the finger guard, and the
air‑flow diverter to the cooling assembly's heatsink
using the four screws and washers.
Make sure that the air‑flow diverter [P] is equally
spaced on the heatsink.
P
Replacing the osmometer module
Refer to
and callouts.
1.
2.
3.
4.
Reconnecting the control PCB
Refer to
callouts.
1.
2.
O
3.
4.
5.
6.
7.
8.
9.
Replacement Procedure
OsmoTECH® PRO Block Probe
Removing the osmometer module
Mount the replacement osmometer module to the
osmometer module mounting bracket using the four
socket head cap screws and washers [F].
Secure the SPG and BPG ring terminals to the side
of the osmometer module mounting bracket using
the hex nut and washer [E].
Slide the osmometer module into the instrument.
Mount the osmometer module mounting bracket to
the chassis using the four socket head cap screws
and washers [D].
Disconnecting the control PCB
Place the instrument in an upright position.
Plug the A1, A2, A3, and A4 connectors on the
cooling system harness into their mating connectors.
Plug the B1, B2, B3, B4, B6, B7, B8, B9, and B10
connectors on the motor/sensor cable harness into
their mating connectors.
Plug the C10 and C11 connectors into the control PCB.
Secure the C10 sample probe cable and the C11
block probe cable to the side of the power supply
and controls assembly with the cable clamp and
screw [A].
Reassemble the instrument by performing the
reassembly instructions in OsmoTECH PRO
Disassembly and Reassembly.
When the instrument powers up, press SETTINGS >
PROBE RESISTANCE from the main menu.
Change the block probe resistance value to match
the value previously recorded.
Calibrate the instrument (see Calibration in the
service guide).
for images
for images and
Page 5 222123PM Rev 0