some time to follow the change of pressure by holding
the pipette tip in its position for at least 2 sec. after the
aspiration and the blow out stroke.
If in extreme cases this method of operation does not
result in accurate values, a compensation could be
achieved as follows:
Weigh the liquid pipetted when the pipette is set to the
nominal value. Then calculate the set-off from the nomi-
nal value:
Corr: val. = 2 x nom. val. -
- weight of the sample
m
γ
- density of liquid
Check this operation once again and correct if neces-
sary. Note the corrected value for further pipetting the
same kind of liquid.

6 - PIPETTE TIPS

A.B.T. tips are made from high performance polypropy-
lene and their quality guarantees the precision and accu-
racy associated with the pipette. Strict control is main-
tained throug hout the manufacturing process to ensure
the highest quality.
The use of inferior quality tips will seriously degrade the
performance of the pipette.
Tips 10
These tips are used for volumes between 0.1 μl and 10 μl.
They are used with the AB2 and AB10 models.
Tips 200
These tips are used for volumes between 2 μl and 200 μl.
They are used with the AB20, AB50, AB100 and AB200
models.
Tips 300
These tips are used for volumes between 50 μl and 250 μl.
They are used with the AB250 models.
Tips 1000
These tips are used for volumes between 100 μl and
1000 μl. They are used with the AB1000 models.
Tips 5000
These tips are used for volumes between 500 μl and
5000 μl. They are used with the AB5000 models.
Tips 10000
7
m
-
γ
These tips are used for volumes between 1000 μl and
10000 μl. They are used with the AB10000 models,
which are equipped with a white pushbutton.

7 - RECOMMENDATIONS

Observing the following recommendations will ensure
maximal possible accuracy and precision of liquid sam-
pling.
• Make sure to operate the pipette slowly and smoothly.
• The depth of immersion in the sample liquid should be
the minimum necessary and should remain constant
during aspiration.
• The pipette should be held in a vertical position.
• Change the tip when volume setting is chan ged or when
a different liquid is to be aspirated.
• Change the tip if a droplet remains on the end of the tip
from the previous pipetting operation.
• Each new tip should be pre-rinsed with the liquid to be
pipetted.
• Liquid should never enter the pipette shaft. To prevent
this:
- Press and release the pushbutton slowly and smoothly.
- Never turn the pipette upside down.
- Never lay the pipette on its side when there is liquid in
the tip.
• Never force the volume setting beyond its recommended
limits.
• When pipetting liquids with temperatures different from
the ambient temperature, it is reco mmended to pre-
rinse the tip several times before use.
• Do not pipette liquids with temperatures above 70°C.
• When pipetting acids or corrosive solutions which emit
vapours, it is recommended to disassemble the shaft
and to rinse the piston and seal with distilled water after
finishing the pipetting operation.

8 - RECALIBRATION

Pipettes are calibrated by gravimetric me thod, using tips
and distilled water, at the temperature 20±1°C, accord-
ing to EN ISO 8655 standard.
If during pipette operation you find that the accuracy
error (the difference between the real aspirated volume
ENGLISH