- ページ 13

医療機器 Droper Field 1000のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Droper Field 1000 15 ページ。 Mechanical pressure infuser
Droper Field 1000 にも: 取扱説明書 (16 ページ)

Re-installing the components
After complete drying, reseat the plate by presenting it, as shown, opposite to
its clamping clips and making a strong pressure movement. Visually check that
the different elements are correctly positioned.

11. PERIODIC INSPECTION

The device is designed to operate more than 6,000 times under normal operating conditions.
However, the manufacturer advises that users perform annual maintenance and inspections to
ensure the correct operation of the device, either by sending it to the manufacturer or distributor, or
by using the method described below :
Place a 1,000 ml bag of NaCl 0.9%, the design of which corresponds to the plate's surface in the
Droper Field and introduce the spike of a classic infusion line without a catheter or needle, and with
the roller clamp in the OFF position. Open the infusion line into a recipient placed at the same height
as the device and measure the time required to drain the bag. The bag must be drained in less than
or equal to 10 minutes.
12. CONDITIONS OF TRANSPORT, STORAGE
AND USE
Storage of the device when not in use should be in a dry and temperate place
Relative humidity : 20% to 90%
Storage, transport and use temperature : - 10°C to + 50°C
13. WARRANTY CONDITIONS
The Droper Field is guaranteed against any part or manufacture defects for a period of 2 years from
its date of first use. To benefit from the parts and labour warranty from our After-Sales Service or
from an approved service, the following conditions must be complied with :
The device must have been used under the normal operating conditions described herein.
The device must not have suffered from any deterioration caused by its storage,
maintenance or incorrect handling.
The device must not have been modified or repaired by persons not authorized by the
manufacturer.
For item returns and repairs, please contact the After-Sales Service
Instructions for use - Version 2
13/15