- ページ 9

医療機器 Beckman Coulter UniCel DxC 600のPDF 毎日のスタートアップをオンラインで閲覧またはダウンロードできます。Beckman Coulter UniCel DxC 600 15 ページ。 Synchron clinical system

Beckman Coulter UniCel DxC 600 毎日のスタートアップ

Unload CC Reagents

Select Rgts/Cal from the Menu bar
Page down to locate the position(s) on the CC carousel
where reagent cartridge(s) will be unloaded (pages 3 to 8)
Select all Positions to
unload reagent cartridges
Select Load F1
When prompted by the software,
open the reagent carousel door
Grasp the cartridge in the load position (located at the
blinking bar code reader) and move it in front of the barcode
reader beam or select Clear F1 to remove reagent
information. Verify the reagent information is removed from
the monitor. Remove the cartridge
Close the reagent carousel door
Additional
Yes
reagents
selected?
No
Procedure complete
UniCel DxC, Customer, Daily Start Up Job Aid
Page 9 of 15
Rev. F (September 2013)
Beckman Coulter Professional Development