- ページ 2
計測機器 Ankom TDF23のPDF サービス手順をオンラインで閲覧またはダウンロードできます。Ankom TDF23 9 ページ。 Paddle bar replacement
3.
Unplug the black 25 pin Cable Connector plug.
Twist counter-clockwise from the underside of the Electrical
Enclosure (right side from the front) and pull to release the
plug.
4.
Unscrew the Shoulder Screw that connects the cable block
to the mixer piston.
Using the 1/8" Hex key wrench, unscrew the Shoulder
Screw on the Paddle Bar Assembly sticking through the hole
in the front of the instrument, connecting to the brass Rod
End of the Mixer Piston. Set the Shoulder Screw aside for
later use.
5.
Unscrew the 4 screws holding the Paddle Bar Assembly
onto the front of the instrument.
a. Lift the instrument to its upright position.
b. Use the 5/32" Hex key wrench to remove the screws.
Be careful while doing this as the Paddle Bar Assembly
will fall out when the last screw is removed.
ANKOM Technology 2052 O'Neil Road Email: [email protected] Phone: 315-986-8090 Fax: 315-986-8091