- ページ 3

ラボラトリー機器 Drucker 614BのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Drucker 614B 6 ページ。 Laboratory centrifuge

WARRANTY:
The Drucker Company
warranties that this centrifuge
is free from defects in
workmanship and parts
for 12 months.
200 Shadylane • Philipsburg, PA 16866
Phone: 814-692-7661 • Fax: 814-692-7662 • www.druckercompany.com
Made in the USA by
Page 10

Features:

• Fixed-angle rotor for spinning 75mm, 100mm and 125mm test tubes in
specially designed tube holders
• Cool–Flow air flow design that prevents overheating of samples
• Heavy gauge steel construction for safety and durability
• Lid safety switch that prevents the centrifuge from operating unless the
lid is closed and latched
• Removable rotor for easy cleaning
• Brushless A/C motor
• Clear lid for safe observation of samples and optical calibration of speed

Specifications:

General Specifications for the Model 614B Centrifuge
Nominal Speed (125mm holders):
Nominal RCF (125mm holders):
Nominal Speed (100mm holders):
Nominal RCF (100mm holders):
Nominal Speed (75mm holders):
Nominal RCF (75mm holders):
Maximum capacity:
Overall Dimensions (H x W x D):
Centrifuge Motor:
Nominal Acceleration Time:
Protection Breaker:
Timer:
Current Requirement:
Voltage Requirement:
Frequency:
Weight:
Before using any cleaning or decontamination methods except those recommended
by the manufacturer, users should check with the manufacturer that the proposed
method will not damage the equipment. See page 7, (bottom), for the recommended
cleaning solutions.
Any use other than those specified by the Manufacturer is explicitly prohibited.
*
Maximum sample density is 1.15 grams / mL,
(water density = 1.0 grams / mL)
3,150 (± 100) RPM
1,200 (± 80) xg
3,250 (± 100) RPM
1,080 (± 80) xg
3,300 (± 100) RPM
950 (± 80) xg
90 mL (6 x 15 mL)*
8.5 in. x 11 in. x 12.5 in.
1/30 HP, AC
45 seconds
4 Amp. re–settable
mechanical, 1 to 30 minutes
accuracy ± 10%
1.0 Amps
115 (±10) Volts
60 Hz
18.5 lbs.
Page 3