- ページ 14
ラボラトリー機器 Drucker Diagnostics 614VETのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Drucker Diagnostics 614VET 16 ページ。
The machine does not unlock
after a run has completed.
The run time is not set to the
desired length.
E
MERGENCY
In the event of power failure, it may be impossible to unlock the lid by
conventional means. In this case, entry into the rotor chamber may be
made by removing the latch label and using a
pen to manually disengage the locking
mechanism (see photo). Pull the mechanism
towards the control panel and then unlatch
and open the lid. If the unit is damaged,
contact your authorized dealer or Drucker
Diagnostics.
C
ALIBRATION
It is recommended that the top speed be tested every two years for
continued safe operation. Contact Drucker Diagnostics for further
information or testing availability.
Drucker Diagnostics +1-814-692-7661 - [email protected] Page | 14
R
C
OTOR
HAMBER
T
ESTING
The lid should remain locked until
the rotor has nearly come to a
complete stop and then unlock
for 60 seconds. If additional
unlock time is needed, press the
'OPEN/STOP' button with
the machine plugged in and the
rotor stopped. If the lid remains
locked after this and will not
unlock, the electronics may have
been damaged. Contact Drucker
Diagnostics for assistance. To
access the rotor chamber, follow
the procedure on page 14,
"Emergency Rotor Chamber
Entry".
Check the run preset by following
the instructions on page 8. If the
preset is not the desired length,
follow the procedure on the same
page to change the run preset
time.
E
NTRY